(Key:
[x]
available in data protocol/documented in spec and implemented[-]
available in data protocol/documented in spec but not yet implemented[?]
unknown whether available in data protocol/documented in spec; not yet implemented*[ ]
TODO: needs implementation![ ]
unavailable in data protocol and/or not documented in spec and not yet implemented)
[x]
cbg values[ ]
units of cbg values (read from device, not hard-coded)[x]
out-of-range values (LO or HI)[x]
out-of-range value thresholds (e.g., often 40 for low and 400 for high on CGMs)
Device-specific? (Add any device-specific notes/additions here.)
- internal glucose unit is always mg/dL for this device, independent of display unit
- out-of-range thresholds are 41 mg/dL and 499 mg/dL
- out-of-range measurements are reported as values 40 or 500 respectively
[ ]
calibrations[ ]
calibration value[ ]
units of calibration value (read from device, not hard-coded)
[x]
time changes (presence of which is also in the BtUTC section below)[x]
device display timefrom
(before change) andto
(result of change)[x]
agent of change (automatic
ormanual
)[ ]
timezone[ ]
reason for change (read from device)
Device-specific? (Add any device-specific notes/additions here.)
- device does not need calibration
[x]
units preference for BG display[x]
units of data being uploaded (will be mutated to mmol/L storage units if not mmol/L)[x]
transmitter ID[ ]
low alert settings[ ]
enabled[ ]
level/threshold[ ]
snooze threshold
[ ]
high alert settings[ ]
enabled[ ]
level/threshold[ ]
snooze threshold
[ ]
rate-of-change alerts[ ]
fall rate alert[ ]
enabled[ ]
rate threshold for alerting
[ ]
rise rate alert[ ]
enabled[ ]
rate threshold for alerting
[ ]
out-of-range alerts[ ]
enabled[ ]
snooze time between alerts
[ ]
predictive alerts[ ]
low prediction[ ]
enabled[ ]
time sensitivity (minutes to predicted low for alerting)
[ ]
high prediction[ ]
enabled[ ]
time sensitivity (minutes to predicted high for alerting)
[ ]
calibration alerts/reminders[ ]
pre-reminder[ ]
overdue alert
Settings history:
[ ]
device stores all changes to settings OR[x]
device only returns current settings at time of upload
No Tidepool data model (yet): volume and/or vibrate mode of all alerts (can/should go in payload
).
Device-specific? (Add any device-specific notes/additions here.)
[x]
index[ ]
UTC timestamp (Hey, one can dream!) OR[x]
internal timestamp or persistent log index (across device communication sessions) to order all pump events (regardless of type), independent of device display time OR[ ]
ephemeral log index (does not persist across device communication sessions) to order all pump events (regardless of type), independent of device display time
[x]
date & time settings changes[x]
usecommon.checkDeviceTime(currentDeviceTime, timezone, cb)
to check against server time
Device-specific? (Add any device-specific notes/additions here.)
NB: You can and should add to this section if there are other data types documented in the device's data protocol specification but not part of Tidepool's data model (yet).
[-]
activity/exercise[-]
food (e.g., Dexcom allows logging carb events)[-]
notes/other events[-]
insulin (rapid acting, long term)
Choose one of the following:
[x]
legacy "jellyfish" ingestion API*[ ]
platform ingestion API
Use this space to describe device-specific known issues or implementation TODOs not contained in the above datatype-specific sections.
[x]
smbg values[ ]
units of smbg values (read from device, not hard-coded)[x]
out-of-range values (LO or HI)[x]
out-of-range value thresholds (e.g., often 20 for low and 600 for high on BGMs)[x]
date & time settings changes[x]
blood ketone values[ ]
units of blood ketone values (read from device, not hard-coded)[x]
ketone out-of-range values[x]
ketone out-of-range value thresholds[x]
usecommon.checkDeviceTime(currentDeviceTime, timezone, cb)
to check against server time
Device-specific? (Add any device-specific notes/additions here.)
- internal glucose unit is always mg/dL for this device, independent of display unit
- glucose out-of-range thresholds are 41 mg/dL and 499 mg/dL
- glucose out-of-range measurements are reported as values 40 or 500 respectively
- ketone out-of-range upper threshold is 8.0 mmol/L
[x]
control (solution) tests (whether marked in UI or auto-detected) - until we have a data model, these should be discarded[-]
device settings, other than date & time (e.g., target blood glucose range)[-]
tag/note (e.g., pre- vs. post-meal)
Choose one of the following:
[x]
legacy "jellyfish" ingestion API*[ ]
platform ingestion API
Use this space to describe device-specific known issues or implementation TODOs not contained in the above datatype-specific sections.